Troubleshooting, The test tool shuts down after a short time, The screen remains black – Fluke Biomedical 190M User Manual

Page 116: The test tool cannot be turned off, Flukeview does not recognize the test tool

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

106

Troubleshooting

The Test Tool Shuts Down After a Short
Time

The batteries may be empty. Check the battery
symbol at the top right of the screen. A

symbol

indicates that the batteries are empty and must be
charged. Connect the BC190 power adapter.

The test tool is still on but the ‘display auto off’ timer is
active, see Chapter 6 ‘Setting the Display AUTO-off
Timer’. To turn the display on press any key (restarts
the ‘display AUTO-off’ timer), or connect the BC190
power adapter.

The power down timer is active, see Chapter 6
‘Setting the Power Down timer’.

Press

to turn the test tool on.

The Screen Remains Black

Make sure that the test tool is on (press

)

You might have a problem with the screen contrast.

Press

USER

, then press

F4

. Now you can use

the arrow keys to adjust the contrast.

The display ‘auto off’ timer is active, see Chapter 6
‘Setting the Display AUTO-off Timer’. To turn the
display on press any key (restarts the ‘display AUTO -
off’ timer), or connect the BC190 power adapter.

The Test Tool Cannot Be Turned Off

If the test tool cannot be turned off due to a software
hangup, do the following:

Keep the ON/OFF key pressed for at least 5 seconds.

FlukeView Does Not Recognize the Test
Tool

Make sure that the test tool is turned on.

Make sure that the interface cable is properly
connected between the test tool and the PC. Use only
the test tool mini USB port for communication with a
computer!

Make sure that no SAVE/RECAL/COPY/MOVE action
from or to a USB-stick is being performed.

Make sure that the USB driver has been correctly
installed, see Appendix A.