background image

Using the Scope and Meter

Acquiring Waveforms

1

31

Comparing Waveforms

You can display a fixed reference waveform with the
actual waveform for comparison.
To create a reference waveform and to display it with the
actual waveform, do the following:

1

SCOPE

Display the

SCOPE

key labels.

2

F4

Open the Waveform Options

menu.

3

ENTER

Jump to the Waveform field and
select Reference… to open the

WAVEFORM REFERENCE

menu.

4

ENTER

Select On to display the reference
waveform. This can be:
- the last used reference waveform

(if not available no reference
waveform will be shown).

- the envelope waveform if the

persistence function Envelope is
on.

Select Recall… to recall a saved
waveform (or waveform envelope)
from memory and use it as a
reference waveform.
Select New… to open the

NEW

REFERENCE

menu.


If you selected New… continue at
step 5, else go to step 6.

5

Select the width of an additional
envelope to be added to the
momentary waveform.

Advertising