Miscellaneous, Display, Power – Fluke Biomedical 190M User Manual

Page 132: Probe calibration, Internal memory

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

122

Miscellaneous

Display

View Area .............. 126.8 x 88.4 mm (4.99 x 3.48 inches)

Resolution ................................................ 320 x 240 pixels

Backlight ...................... LED (Temperature compensated)

Brightness ................................ Power Adapter: 200 cd/m

2

Battery Power: 90 cd/ m

2

Display Auto-OFF time (battery saving) ......... 30 seconds,

5 minutes or disabled

Power

For Model190M-4:
Rechargeable Li-ion Battery (model BP291)
Operating Time ................. up to 7 hours (Low Intensity)
Charging

Time

....................................................

5

hours

Capacity/Voltage

..................................... 52 Wh / 10.8 V


For Model 190M-2:
Rechargeable Li-ion Battery (model BP290):
Operating Time ................. up to 4 hours (Low Intensity)
Charging Time ................................................. 2.5 hours
Capacity/Voltage

..................................... 26 Wh / 10.8 V

Rechargeable Li-ion Battery (model BP 290 and BP291):
Lifetime (> 80 % capacity) .......... 300x charge/discharge

Allowable

ambient

temperature during charging: .... 0 to 40

°

C (32 to 104

°

F)

Auto power down
time (battery saving): .............. 5 min, 30 min or disabled

Power Adapter: BC190/808 Universal Switchable Adapter
115 V ±10 % or 230 V ±10 %, with plug EN60320-2.2G

Line Frequency ............................................ 50 and 60 Hz

Probe Calibration

Manual pulse adjustment and automatic DC adjustment
with probe check

Generator Output ................................1.225 Vpp / 500 Hz

square

wave

Internal Memory

Number of Scope Memories ..................................... 15/30
Each memory can contain 2/4 waveforms plus
corresponding

setups

Number of Recorder Memories ................................... 2/10
Each

memory

can

contain:

a 2/4 channel input TrendPlot

a 2/4 channel input Scope Record

100 2/4 channel input Scope screens (Replay)

Number of Screen Image memories ............................. 1/9
Each memory can contain one screen image

Advertising