PD6730X Vantageview Super Snooper Modbus Scanner

Instruction Manual

12

Pulse Rate/Totalizer

DISPLAY
ASSIGNMENT

The top display may be assigned to rate, total, or grand total, in addition to
Modbus process variables.

RATE DISPLAY
UNITS

Gallons, liters, imperial gallons, cubic meters, barrels, bushels, cubic yards,
cubic feet, cubic inches, liquid barrels, beer barrels, hectoliters, or custom.

RATE DISPLAY TIME
BASE

Rate display may be calculated in terms of units per second, minute, hour, or
day.

TOTAL & GRAND
TOTAL DISPLAY
UNITS

Gallons, liters, imperial gallons, cubic meters, barrels, bushels, cubic yards,
cubic feet, cubic inches, liquid barrels, beer barrels, hectoliters, or custom.
Setting is independent for each.

TOTAL & GRAND
TOTAL DISPLAY
UNIT MULTIPLIER

x1, x100 (h), x1000 (k), or x1,000,000 (M) multiplier (and prefix) applied to total
or grand total display units. Setting is independent for each.

TOTAL & GRAND
TOTAL DECIMAL
POINT

Up to six decimal places or none:
6.666666, 55.55555, 444.4444, 3333.333, 22222.22, 111111.1 or 0000000
Total and grand total decimal points are independently programmed, and are
independent of rate decimal point.

TOTALIZERS

Calculates total and grand total based on rate and field programmable multiplier
to display total in engineering units. Time base must be selected according to the
time units in which the rate is displayed. The total and grand total utilize the
same time base, with different conversion factors and resets.

TOTALIZER
RESET

Via SafeTouch

®

RESET button, mechanical button (cover off), external contact

closure (total only), automatically via user selectable preset value and time delay
(1 – 99,999 sec). Manual reset may be disabled or protected by password for the
total and grand total. Total and grand total reset independently.

TOTAL
OVERFLOW AND
ROLLOVER

The total can display up to 9,999,999,999,999. Up to 9,999,999 can be displayed
on the lower display normally. An overflow display will toggle between the first
six digits and last seven digits (999999 <> 9999999) for a 13-digit total. The total
will rollover beyond thirteen digits. The T indicator on the display will flash to
indicate total overflow, and the six most significant digits (first six numbers of the
total) are indicated with the flashing overflow symbol .

GRAND TOTAL
OVERFLOW AND
ROLLOVER

The grand total can display up to 9,999,999,999,999. Up to 9,999,999 can be
displayed on the lower display normally. An overflow display will toggle between
the first six digits and last seven digits (999999 <> 9999999) for a 13-digit total.
The grand total will rollover beyond thirteen digits. The GT indicator on the
display will flash to indicate grand total overflow, and the six most significant
digits (first six numbers of the grand total) are indicated with the flashing overflow
symbol .

EXTERNAL
TOTAL RESET

External total reset connections are made between RST and COM. Logic High:
1.4 V, 3.3V max; Logic Low: < 0.8 V.
90 ms minimum pulse width.