Communications interface 15.2 – Omega RD200 User Manual

Page 104

Advertising
background image

-100-

15.

COMMUNICATIONS INTERFACE 15.2

Programming communications


This explanation is only for the communications interface (option). Press and simultaneously for 3
seconds or more to display the “communications” programming screen.

Operation screen

[

Checking communications type

]

(

+ )

3 sec. or more

 

RS232C?

[Programming address]

to

  

Temporarily stored

[Selecting transmission speed]

  

Temporarily stored

[Selecting characters]

  

Temporarily stored

[Selecting checksum]

  

Temporarily stored


Stored

Communications type is displayed. The
communications type cannot be changed

.




01 to 99:  (private) protocol
01 to 31:  (MODBUS) protocol
(Note) “01” is displayed only when “private”

is changed to “MODBUS”.



1200, 2400, 4800, 9600,19200 bps
Only 9600 and 19200 bps are available
with the MODOBUS protocol.
(Note) “9600” is displayed only when
“private” is changed to “MODBUS”.



Characters are displayed with codes.

Code

Character

length

Parity Stop

bit

7E1

7 bits

Even

1

7E2

7 bits

Even

2

701 7

bits Odd

1

702 7

bits Odd

2

8N1

8 bits

Non

1

8N2

8 bits

Non

2

8E1

8 bits

Even

1

8E2

8 bits

Even

2

801 8

bits Odd

1

802 8

bits Odd

2

(Note) “8N1” is displayed only when
“private” is changed to “MODBUS”.



 

: Checksum ON
: Checksum OFF
(Note) No display appears for the

MODBUS protocol.



Pressing for each programming
executes temporary storing. Pressing
for 2 seconds or more executes storing
and returns to the operation screen.


If the screen returns to the operation
screen before pressing , the
parameter is returned to the parameter
before programming.

NO

Address

Transmission speed

Characters

2 sec. or more

2 sec. or
more

A

2 sec. or more

Communications type check

Address programming range

Transmission speed selection
range

Character selection

Checksum selection range

From “ temporarily stored” to

“stored”

Returning to operation screen

SHIFT

(-)

DISP

SHIFT

(-)

YES

ENTRY

SET

END

0

CLEAR

9

LIST

ENTRY

DISP

%

/

ENTRY

DISP

%

/

ENTRY

DISP

%

/

Checksum

ENTRY

DISP

ENTRY

DISP

ENTRY

Reference 1

Reference 2

Reference 3

Reference 4

Reference 5

Reference 6

Reference 7

A

A

A

A

A

2 sec. or
more

2 sec. or more

Ref. 5

Ref. 6

Ref. 4

Ref. 3

Ref. 2

Ref. 1

Ref. 7