22. SPECIFICATIONS

9

Accuracy Ratings

No. Input type

Measuring range

Reference

range

Accuracy

rating

01

-13.80 to 13.80mV

±13.8mV

02

-27.60 to 27.60mV

±27.6mV

03

-69.00 to 69.00mV

±69.0mV

04

-200.0 to 200.0mV

±200.0mV

05

DC

(mV)

-500.0 to 500.0mV

±500.0mV

06

-2.000 to 2.000V

±2V

07

-5.000 to 5.000V

±5V

08

-10.00 to 10.00V

±10V

09

-20.00 to 20.00V

±20V

10

DC

(V)

-50.00 to 50.00V

±50V

±0.1% ± 1 d

21

-200.0 to 300.0°C

±13.8mV

22

-200.0 to 600.0°C

±27.6mV

23

K

-200.0 to 1370 °C

±69.0mV

24

-200.0 to 200.0°C

±13.8mV

25

-200.0 to 350.0°C

±27.6mV

26

E

-200 to 900°C

±69.0mV

27

-200.0 to 250.0°C

±13.8mV

28

-200.0 to 500.0°C

±27.6mV

29

J

-200 to 1200°C

±69.0mV

30

-200.0 to 250.0°C

±13.8mV

31

T

-200.0 to 400.0°C

±27.6mV

32

0 to 1200 °C

±13.8mV

33

R

0 to 1760 °C

±27.6mV

34

0 to 1300 °C

±13.8mV

35

S

0 to 1760 °C

±27.6mV

36

B

0 to 1820 °C

±13.8mV

±0.1% ± 1 d

37

-200.0 to 400.0°C

±13.8mV

38

-200.0 to 750.0°C

±27.6mV

39

N

-200 to 1300 °C

±69.0mV

40 W・WRe26

0 to 2315°C

±69.0mV

±0.15% ± 1 d

41

WRe5・

WRe26

0 to 2315°C

±69.0mV

43

PtRh40-

PtRh20

0 to 1888°C

±13.8mV

44

-50.0 to 290.0°C

±13.8mV

45

-50.0 to 600.0°C

±27.6mV

46

NiMo-Ni

-50 to 1310 °C

±69.0mV

47 CR-AuFe

0.0 to 280.0K

±13.8mV

±0.2% ± 1 d

48 0.0

to

350.0°C

±13.8mV

49 0.0

to

650.0°C

±27.6mV

50

Platinel ll

0 to 1395°C

±69.0mV

51

-200.0 to 250.0°C

±13.8mV

52

-200.0 to 500.0°C

±27.6mV

53

U

-200.0 to 600.0°C

±69.0mV

±0.15% ± 1 d

54

-200.0 to 250.0°C

±13.8mV

55

-200.0 to 500.0°C

±27.6mV

56

L

-200 to 900°C

±69.0mV

±0.1% ± 1 d

70 -140.0

to150.0°C

160

±0.15% ± 1 d

71

-200.0 to 300.0°C

220

72

Pt100

(JIS’97)

-200.0 to 850.0°C

400

±0.1% ± 1 d

73

-140.0 to 150.0°C

160

±0.15% ± 1 d

74

-200.0 to 300.0°C

220

75

Pt100

QPt100

JIS’89

-200.0 to 649.0°C

400

±0.1% ± 1 d

76

-140.0 to 150.0°C

160

±0.15% ± 1 d

77

-200.0 to 300.0°C

220

78

JPt100

-200.0 to 649.0°C

400

±0.1% ± 1 d



No. Input type

Measuring range

Reference

range

Accuracy

rating

79

Pt50

-200.0 to 649.0°C

220

±0.1% ± 1 d

80

Pt-Co

4.0 to 374.0K

220

±0.15% ± 1 d

Note) “

± 1 d” at accuracy rating stands for ± 1 digit.

Note) Thermocouple inputs do not include reference junction

compensation accuracy.
The Indication equivalent to 200

µV may vary under the

test environment of EMC directive.

Exceptions to accuracy rating

Input type

Measuring range

Accuracy rating

K, E, J, T, L

-200 to 0°C

R, S

0 to 400°C

±0.2% ± 1 digit

0 to 400°C

Not specified

B

400 to 800°C

±0.15% ± 1 digit

N, U

-200 to 0°C

±0.3% ± 1 digit

0 to 100°C

±4% ± 1 digit

W-WRe26

100 to 300°C

±0.5 % ± 1 digit

0 to 300°C

±1.5% ± 1 digit

PtRh40-PtRh20

300 to 800°C

±0.8% ± 1 digit

0 to 20K

±0.5% ± 1 digit

CR-AuFe

20 to 50K

±0.3% ± 1 digit

Pt100 (’97)

700 to 850°C

±0.15% ± 1 digit

Pt-Co

4 to 50K

±0.3% ± 1 digit

Note) Thermocouple input is converted into the reference