Programming, 3 list of programming items – Omega RD200 User Manual

Page 39

Advertising
background image

-35-

7. PROGRAMMING

7.3 List of Programming Items

Parameters

Key operations

Programming

items

Defaults

Programming ranges

Ref.

page

Alarm dead

band

+

(3 sec. or more)

0.1

0.1 to 9.9 (%): % programming of scale
span

11.6

Start time

Not

programmed

00:00 to 23:59

Periodic

data printing

+

Interval time

Not

programmed

00H05* to 23H59 *Limited by chart speed.

11.7

Engineering

unit *

+

V

Max. 5 digits
(Including numeric values, alphabetic
characters, %, /,

°C, °F and space)

Reflected by “Range No.” programmed
(mV, V,

°C, k)

11.8

Tag *

+

Not

programmed

Max. 9 digits (Including numeric values,
alphabetic characters, %, /,

°C, °F and

space)

11.9

No.

Not

programmed

1 to 5, Five kinds of messages can be
programmed.

Message

+

(3 sec. or more)

Message

Not

programmed

Max. 15 digits
(Including numeric values, alphabetic
characters, %, /,

°C, °F and space)

11.10

Burnout *

+

(3 sec. or more)

non

non, UP bUrn, doWn bUrn
For channels with voltage range, burnout
is disabled (“non”), even if burnout is
programmed (up or down).

11.11

Pass code

non

non (not programmed). For programming,
use 4 digits from 0 to 9.

Passcode/

keylock

(3 sec. or more)

Key lock

UnLoCkEd

LockEd, UnLoCkEd

11.12

Time axis

synchronization

(3 sec. or more)

oFF on,

oFF

9.6

Input filter*

(3 sec. or more)

Time

constant

0

0, 0 to 10 (sec.). “0” means no time
constant.

11.13

Copy

Programming in each
programming mode

Programmed parameters can be copied to
other channels. Available parameters to be
copied are [Range/Printing range],
[Engineering unit] and [Tag].

11.14

* Programming is required to each channel. For 1-pen type, programming is only for CH 1. The items without *

mark are programming common to all channels. For "alarm", programming is required to each level of
channels.

3

Programming for Options

Parameters

Programming

items

Key operations

Programming

contents

Defaults

Programming ranges

Ref.

page

Relay

No.

+ Relay

No.

0

"0" to "n" (n: 6, 12)*
In alarm programming mode,
program relay Nos. (terminal
Nos.) for alarm output.
* number of output points

AND/

OR

(3 sec. or more)

AND/OR or

AND/OR programming is required
for each relay No.

Relay coil

E

E (energize), d (not energize)

Relay output

U

H (hold), U (not hold)

Alarm

output

Output

mode

(3 sec. or more)

Alarm display

U

H (hold), U (not hold)

Common

to all relay

Nos.

12.1

SHIFT

SHIFT

SHIFT

SHIFT

0

CLEAR

0

CLEAR

0

CLEAR

5

DATA I

3

UNIT

6

A - Z

/

9

LIST

8

ALARM

°C

SHIFT

SPACE

COPY(=)

SHIFT

8

ALARM

DISP

°C

°C

0

CLEAR

SHIFT

6

A - Z

SHIFT

%

Advertising
This manual is related to the following products: