-104-

18. ADJUSTMENT

18.1 Adjustment of Measured Values

Adjustment

Adjustment comprises four kinds shown below. (1), (3) and (4) have already been adjusted. However, it is
recommended for maintaining the measuring and printing accuracy to adjust them once a year.

Calibration Details

Reference

(1) Measured values adjustment

Adjustment to keep the measured value against input within
the accuracy ratings.

Sec.18.1

(2) Shift programming of

measured value

Programming to shift a measured value

Sec.18.2

(3) Adjustment of printing position

Adjustment to set the printing range to the zero and span lines
on the chart

Sec.18.3

(4) Adjustment for time-axis

synchronization of pens

Adjustment of the gaps between pens for the time-axis
synchronization

Sec.18.4

1

Adjustment of measured values

It is recommended for maintaining measuring and
printing accuracy to adjust them once a year.
(1) Execute the adjustment for each channel.
(2) Execute the adjustment under the reference

condition. (See the right table.)


Items Reference

conditions

Ambient temperature 23

±2°C

Ambient humidity

55

±10%RH

Power voltage

100

±1VAC

Power supply

frequency

50 or 60

±0.5Hz

2

Preparation

1) Preparation of tools

Input types

Tools

DC voltage Thermocouple

Resistance

thermometer

Remarks

DC standard

voltage/current generator

Accuracy: Should be better than

±0.05%.

Reference junction

compensator

0

°C ±0.2°C or less

Thermocouple for test

Same type of thermocouple as input type

Standard variable resistor

Accuracy: Should be better than

±0.05%.

3-core copper wire

Same resistance for each 3 cores

2) Connection

Connections depend upon the input types. See the next page.

3) Before starting adjustment

(1) After the connection, mount the terminal board cover and turn on the power supply.
(2) Before starting adjustments, warm up your recorder for at least 30 minutes until it becomes stable. (It is

recommended to warm it up for at least one hour if possible.)


Check and adjustments of the measured values need meticulous work in addition to the standard tools and
reference conditions.
If you need checks or adjustments of the measured values, please consult your nearest CHINO’s agent.

Reference conditions

Adjustments

Remarks