21. MAINTENANCE

21.3 Measured Values Check

3

Connections

Connections depend upon the input types. Connect both standard and other tools to the measuring input
terminals to be adjusted.



In order to prevent electric shock, turn off the power source before starting connections.


(1) In case of thermocouple input

The electromotive force of the thermocouple input is reduced by an amount proportional to the temperature
at the terminals. The recorder itself compensates (reference junction compensation) for this reduction. The
adjustment is executed by inputs based on the reference electromotive force (0

°C reference). In case the

“RJ selection” is set to “1” (Enable) in the “Range/Printing range” programming, an amount equivalent to the
reference junction compensation has to be subtracted using a reference junction compensator.

(2) DC voltage input

(3) Resistance thermometer input

4

Checking Method

Input the printing range (T) of 0%, 50% and 100% from a
standard tool and obtain an error (e) by reading the
measured values (M).



e (%) = X 100

Error computation

(+)

(-)

(B)

(A)

(B)

*Thermocouple of the same
type as the input type

Input terminals

1 2 3 4

( - )

(+)

Thermocouple wire *

Copper lead wire

Test tube

Vacuum bottle

Ice + distilled water

Silicon oil

Reference junction

compensator

DC standard voltage

generator

Standard variable

resistor

Input terminals

1

2 3 4

Input terminals

1

2 3

4

Standard variable

resistor

M – T

T

Turn off the power source before starting connections

Caution