11. OTHER PROGRAMMING

11.4 Subtract Printing


This programming is for printing a difference between channels or between a channel and a reference value.
Press and simultaneously to display the [Range/Printing range] programming screen to program
the subtract printing.

1

Subtract Printing Types

Two types of subtract printings are provided.
(1) Printing a difference between a reference
channel and a subtraction channel
(2) Printing a difference between a reference
channel and a specified subtraction value
(reference value).


Be sure to program [Range/Printing range] of
both the reference channel and the
subtraction channel in advance.


Program the reference value with the scale
programmed value within 5 digits. For
identifying the reference value to a channel
No., be sure to place a decimal point to the
reference value having no decimal point.

<Ex. 2

→ 2.0>




Subtract printing can be specified to any channel.
For example, [CH 1 minus CH 2] can be specified
to [CH 2]. In this case, the difference between
channels are displayed and printed at CH 2.

<Example> Printing the difference between CH 1
and CH 2 at CH 3



2

Programming Modes

1) Mode 1

  

2) Mode 2 (To move from Mode 1 to Mode 2, press after completing

the programming for the Mode 1.)

 



Subtract printing channel

Concept of subtract printing

0

500ºC

0

500ºC

300

200

100

-250

250ºC

Printing range

CH 1

CH 2

CH 3

Equa

l

Ref

e

renc

e

ch

annel

Cu

rso

r

Char

a

ct

e

r

di

sp

la

y

Min

u

s

Unused digits are filled with spaces.

Subtraction channel or
reference value (within 5
digits)

Subt

rac

t

pri

n

ting

Subtract printing range

(11 digits by left justify)

Programming lamp

Programming reference channel
and subtraction channel

Remarks 1

Place a decimal point to the
reference value

SHIFT

7

RANGE

ENTRY

SHIFT

(-)

Remarks 2

Memory
colon

Character display

Programming lamp

“R” stands for
the subtract
printing mode 1

.

“D” stands for
the subtract
printing mode 2

.

Memory
colon

Character display

Char

a

ct

e

r

di

sp

la

y

Unused digits are filled with spaces.

Min. value

(within 5 digits)

to

Max. value

(within 5 digits)

Scale value to

print 0%

+

Scale value to

print 100%

Advertising