4. CONNECTIONS

4.7 Communications terminals

3

RS-422A, RS-485 Connections

RS-422/485 communications interface is connected to a personal computer via a line converter (our Model
SC8-10: sold separately). Three signals of SD, RD and SG are used between the line converter and the
personal computer but any control signal is not used. Wiring processing for control signal in a connector is
necessary in the same as

2

RS-232C connections.

1) RS-422A
















*For the details of the wiring, refer to

2

RS-232C connections.

2) RS-485
















*For the details of the wiring, refer to

2

RS-232C connections”.

Personal computer

(9-pin or 25-pin connector)

*

RD

SD

ER

SG

DR

RS

CS

2

3

4

5

6

7

8

SDA

SDB

RDA

RDB

SG

RD

SD

SG

RDA

RDB

SDA

SDB

1

2

3

4

5

6

7

8

SDA

SDB

RDA

RDB

SG

SDA

SDB

RDA

RDB

SG

Line converter

(SC8-10)

Within 15 m

Within a total length of 1.2 km (max. 31 recorders)

Termination

resistance

100

Ω, 1/4W

Example of 9-in
connector

Personal computer

(9-pin or 25-pin connector)

*

SA

SB

SG

RD

SD

SG

RDA

RDB

SDA

SDB

1

2

3

4

5

6

7

8

SA

SB

SG

SA

SB

SG

Line converter

(SC8-10)

Within 15 m

Within a total length of 1.2 km (max. 31 recorders)

Termination

resistance

100

Ω, 1/4W

Example of 9-in

connector

RD

SD

ER

SG

DR

RS

CS

2

3

4

5

6

7

8

(1)

(2)

(3)

RD200/RD2800 hybrid recorders

(1)

(2)

(3)

RD200/RD2800 hybrid recorders