7. PROGRAMMING

7.3 List of Programming Items

Names

Programming

items

Key operations

Programming

contents

Defaults

Programming ranges

Ref.

page

Speed No.

1

1 to 3

Chart speed
(3 speeds)

Speed

20/25*

0001 to 0600 mm/h or 0001 to 0200
mm/min

13.3

Terminal No.

1

1 to 4

Terminal
allocation

(3 sec. or more)

Operation

non

Selection from 13 items;
non, printing on/off & 3 kinds of chart
speed, message printing (No. 1 to 2),
message printing (No. 1 to 5), operation
record (A/B/C/D), digital data printing, list
printing (No. 1to 3) and totalizing reset

13.2

Printing No.

A

A, B, C, D

Remote

contacts

Operation
record

(3 sec. or more)

Printing
position

0

10 to 90%.
Printing position when the contact is
open. The printing position is at +5 mm
when it is shorted.

13.4

Format

Sd

Select Ar from Sd (standard), Ar, SP, and
PL.

Channel

Not

programmed

1 to 4

Zero for range 1

Not

programmed

a (Minimum value

≤ a < b)

Span for range 1

Not

programmed

b (a < b < c)

Span for range 2

Not

programmed

c (b < c < d)

Span for range 3

Not

programmed

d (c < d < e)

Span for range 4

Not

programmed

e (d < e < f)

Automatic

range shift

(Ar)

Span for range 5

Not

programmed

f (e < f

≤ maximum value)

14.1

Format

Sd

Select SP from Sd (standard), Ar, SP, and
PL.

Channel

Not

programmed

1 to 4

Printing position

for broken point 1

Not

programmed

0 to a (%)

Printing position

for broken point 2

Not

programmed

a to 100 (%)

Scale at zero

Not

programmed

b (Minimum value

≤ b < c)

Scale for broken

point 1

Not

programmed

c (b < c < d)

Scale for broken

point 2

Not

programmed

d (c < d < e)

Compressed/

Expanded

(SP)

Scale for span

Not

programmed

e (d <e

≤ maximum value)

14.2

Format

Sd

Select PL from Sd (standard), Ar, SP, and
PL.

CH for area 1

Not

programmed

1 to 4

CH for area 2

Not

programmed

1 to 4 (Except channels for
area 1)

CH for area 3

Not

programmed

1 to 4 (Except channels for
area 1 and 2)

Record

format

Zone scale

(PL)

CH for area 4

Not

programmed

1 to 4 (Except channels for
area 1 to 3)

Area 1

and 2

only for

RD200

14.3

Protocol

(3 sec. or more)

15.1

Communi-

cations

interface

Specifications

(3 sec. or more)

For details, refer to the separate instruction manual for
“Communications Interface”.

15.2

* RD200: 20 mm/h, RD2800: 25 mm/h

SHIFT

0

CLEAR

0

CLEAR

5

DATA I

2

TAG

1

CHART

SHIFT

SHIFT

SHIFT

DISP

DISP

DISP

SHIFT

SHIFT

(-)

SPACE