21. MAINTENANCE

21.3 Measured Values Check


It is recommended for check measured values once a year for maintaining the measuring and printing
accuracy.

1

Channels to be checked

Check the measured value for each channel.
Different errors may occur for different channels
even in the same range.


To check a channel for which a math-function
or totalization (optional) has been
programmed, program the math expression to
“” before starting the checking.










(Note) Add

±0.01% per 1°C for a case other than

23

±2°C.

2

Preparation

1) Preparation of tools

Input types

Tools

DC voltage

Thermo-coupl

e

Resistance

thermometer

Remarks

DC standard

voltage/current generator

{

{

Accuracy: Should be better
than

±0.05%.

Reference junction

compensator

{

0

°C ±0.2°C

Thermocouple for test

{

Same type of thermocouple as
input type

Standard variable resistor

{

Accuracy: Should be better
than

±0.05%.

3-core copper wire

{

Three copper wires shall have
the same resistance values.


2) Connection

Connections depend upon the input types. See the next page.

3) Before starting adjustments
(1) Mount the terminal board cover and turn on the power supply.
(2) Before starting adjustments, warm up your recorder for at least 30 minutes until it becomes stable. (It is

recommended to warm up for at least one hour if possible.)



Check and adjustments of the measured values need meticulous work in addition to the standard tools and
reference conditions.
If you need checks or adjustments of the measured values, please consult your nearest CHINO’s agent.

Totalization channel

Remarks

Items Reference

conditions

Ambient

temperature

23

±2°C

Ambient humidity

55

±10%RH

Power voltage

100 V AC

±1%

Power supply

frequency

50 or 60 Hz

±0.5 Hz


Reference conditions

Checking

Remarks