12. ALARM OUTPUT

12.4 Programming Output Mode


This explanation is only for the alarm output (option). Press and simultaneously for 3 seconds or

more to display the “Output mode” programming screen. Two output modes, 1) relay coil (energize/not energize),

2) latched alarm display/relay (hold/not hold), are available. The programming is common to all relay Nos..

1

Relay Coil Energize/Not energize

The default is “Energize”.
The terminal configuration differs depending on the type of relay.


1) MOS relay and mechanical relay “a” contact alarm output

Phase

Power off

Alarm reset

Alarm activation

Energized

()

Not

energize

()

2) Mechanical relay “c” contact alarm output

Phase

Power off

Alarm reset

Alarm activation

Energize

()

Not

energize

()

N. O

COM

N. O

COM

N. O

COM

N. O

COM

N. O

COM

N. O

COM

N. O

COM

N. C

N. O

COM

N. C

N. O

COM

N. C

N. O

COM

N. C

N. O

COM

N. C

N. O

COM

N. C

DISP

°C