Adjustment, 1 adjustment of measured values – Omega RD200 User Manual

Page 109

Advertising
background image

-105-

18. ADJUSTMENT

18.1 Adjustment of Measured Values

3

Connections

Connections depend upon the input types. Connect the standard tools to the input terminals to be adjusted.



To prevent any injury caused by electric shock, be sure to turn off the power supply before connections.





(1) In case of thermocouple input

The electromotive force of the thermocouple input is reduced by an amount proportional to the temperature
at the terminals. The recorder itself compensates (reference junction compensation) for this reduction. The
adjustment is executed by inputs based on the reference electromotive force (0

°C reference). In case the

“RJ selection” is set to “1” (Enable) in the “Range/Printing range” programming, an amount equivalent to the
reference junction compensation has to be subtracted using a reference junction compensator.

(2) DC voltage input

(3) Resistance thermometer input

Make sure to connect after turning off the power supply.

Caution

Thermocouple wire *

Copper lead wire

Reference junction

compensator

DC standard voltage

generator

Test tube

Vacuum bottle

Ice + distilled water

Silicon oil

DC standard voltage

generator

Standard variable

resistor

(+)

( - )

+

-

(+)

( - )

+

-

(B)

(A)

(B)

Input terminals

1 2 3 4

Input terminals

1

2 3

4

Input terminals

1

2 3 4

Advertising