10. OPERATIONS

10.2 Operations at Abnormal Inputs

1

Over-range input

The following table shows displays and printings for
input exceeding a printing range or a measuring
range.

• Measuring range:
Determined by the range No. programmed in
“Range/Printing range”. See Section 8.2.
• Printing range:
Trace printing range programmed in “Range/Printing range”.

Display Printing

区分

Input

Digital Bar

graph Digital Analog

(1)

Lower than

minimum

measuring range*

_ _ _ _ _

□□□……□□□

(All go off.)

- - - - -

(2)

Lower than

minimum printing

range

Normal

□□□……□□□

(All go off.)

Normal

Overshooting to

minimum limit

(3)

Higher than

maximum printing

range

Normal

■■■……■■■

(All light.)

Normal

(4)

Higher than

maximum

measuring range*

_ _ _ _ _


■■■……■■■

(All light.)

+ - - - -

Overshooting to

maximum limit

* For the digital display and printing, the measured values can be displayed and printed up to about

±10% of the

span even when the values are outside the measuring range.

2

Input Disconnection

The display and printing, when the input is disconnected, differ depending on the “Burnout enable/disable”
programming. However, for channels of a voltage (mV or V) range, the burnout function is “disable (non)” even
when “Burnout Enable/Disable” is programmed. Program “Burnout enable/disable” to each channel. See
Section 11.11.

Display Printing

Programming burnout

Digital Bar

graph Digital Analog

Disable

(non)

Undefined Undefined Undefined Undefined

Enable - downscale

(doWn bUrn)



□□□……□□□

(All go off.)

BURN

Overshooting to

minimum limit

Enable - upscale

(UP bUrn)



■■■……■■■

(All lights.)

BURN

Overshooting to

maximum limit

Definition of input

Printing range

Measuring range

(1)

(2)

(3) (4)