2. INSTALLATION

2.2

Installation to a Panel



Except portable types, your recorder is designed to install to a panel. Use a panel made of a steel plate of 2
mm to 6 mm in thickness.

RD200

Panel cutout size










Installation

(1) Insert your RD200 (pen type) into the panel

cutout from the front of the panel.

(2) Fix your RD200 (pen type) to the panel by the

mounting brackets (screw tightening torque: 1.0
N•m). Attach 2 mounting brackets to the top and
bottom of this instrument.

RD2800

Panel cutout size











Mounting method

(1) Insert your RD2800 (pen type) into the panel

cutout from the front of the panel.

(2) Gently screw a mounting screw into the screw

hole of the case (left, right).

(3) Attach the mounting bracket to the case by

putting the mounting screw in the large hole of
the bracket. Slide the bracket downward to
attach it closely to the panel and then tighten
the screw with a wrench (screw tightening
torque: 1.2 N•m).

Install your recorder to a panel.

360

36

0

281

+1
0

281

+1

0

(mm)

• Minimum clearance for

closed- installation

200

200

138

+1
0

138

+1

0

(mm)

Mounting bracket

Mounting bracket

Panel thickness

(2 to 6 mm)

Mounting screw

Mounting bracket: Slide it
downward and tighten
with a wrench.

Panel

Mounting bracket

Panel thickness (2 to 6 mm)