Adjustment, 4 time-axis adjustment of pens – Omega RD200 User Manual

Page 113

Advertising
background image

-109-

18. ADJUSTMENT

18.4 Time-axis Adjustment of Pens


When the time-axis synchronization (POC) is programmed to “on”, the gaps between the pens may change over
time, resulting in error between their time-axis. This adjustment is for correction of these changes. It is
recommended to adjust them once a year for maintaining the time-axis synchronization.

Adjustment Flow Chart

Operation screen

[Engineering mode selection]

( + )

3 sec. or more

   

[Selecting adjustment]


   


[Selecting time-axis]


    


[Selecting an adjusting pen]

     

(Starting adjustment )

When (1) the reference pen prints a straight line on
0% to 100% of the chart and (2) Press , the
chart is fed and then the printing of the adjusting pen
coincides with the straight line (Note)
Note) If you press while it does not coincide

with the straight line, re-execute the procedure
from the beginning of [Selecting an adjusting
pen].

(Computation of
correction data)

Temporarily storing

Programming other channel?

Stored











Pens stand by at 0% position of scale.



: Memory clear

: Adjustment

: Hardware check




: Measured value adjustment

: Shift programming for

measured value

: Printing position adjustment

: Time-axis adjustment for pens



Display

Reference

pen

Pen to be

adjusted

(1) 1 P

Plotter pen

1st pen

(2) 2 P

1st pen

2nd pen

(3) 3 P

1st pen

3rdt pen

(4) 4 P

1st pen

4th pen




Store the computed correction data
into memory.

Cursor Adjustment

Cursor

Time axis

2 sec. or more

Cursor

Other pens than the pen being adjusted

Adjusting pen selection

DISP

SHIFT

6

A - Z

°C

°F

ENTRY

°C

°F

ENTRY

°C

°F

(1)

(2) (3)

(4)

ENTRY

ENTRY

ENTRY

(1) Reference pen

prints a straight line.

(2) Initial position of the pen adjusted

Adjustment procedure

(3) When these two
lines match, press
.

ENTRY

Ref. 1

Ref. 2

Ref. 3

Ref. 4

Engineering mode types

Reference 1

Adjustment types

Reference 2

Reference 3

Storing

Reference 4

/

YES

NO

SHIFT

SET

END

Advertising