TROUBLESHOOTING GUIDE

MILLPWR

®

Operation Manual

9-6

System powers up but
motor(s) will not move
table (continued)

Table run away

Table will move in only one
direction.

Should the above recommen-
dation not correct the problem
you are experiencing, contact
your ACU-RITE distributor
for repair and/or replacement
procedures.

Message bar reads: “

(X

OR

Y)

SERVO UNPLUGGED FAULT

.

” Check

the X- and Y-axes servo con-
nections located on the rear of
the Operator Console.

Check the power connection at
each motor and at the Servo
Power connector on the rear of
the Operator Console.

Enable the Motor Monitor
(refer to Diagnostics, Section 7).

Check that the cables are
properly connected to the
correct axes.

Move the table within the
table travel limits. You may
want to consider adjusting
your table travel limits. If so,
refer to Setup).

The scale(s) miscount in
certain areas only (contin-
ued)

Improper X- and/or Y-axes
servo connection.

Motors are not receiving
power

The X- and Y-axes servo
connections located on the
rear of the Operator Console
are reversed.

Table is beyond table travel
limits

Symptom

Recommended
Corrective Action

Probable Cause