4-66

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Vital Sign and ST Segment Trends

Vital Sign and ST Segment Trends

Intended Use

The trends feature of the LIFEPAK 15 monitor/defibrillator provides the ability to graphically

display and document the patient’s vital signs (VS) and ST segment measurements for up to eight

hours. VS trending is intended for use with any patient who requires continuous monitoring of vital

signs over an extended period of time to identify changes in patient condition and to document

patient response to therapy. ST trending is intended for use with patients suspected of having acute

ischemic events, such as unstable angina, and for patients during treatment of an acute ischemic

event. ST segment measurement is initiated using a 12-lead ECG and is derived using the

University of Glasgow 12-Lead ECG Analysis Program.

VS and ST Trends Warning

WARNING

INACCURATE INTERPRETATION OF PATIENT STATUS

Vital sign and ST graphs are tools to be used in addition to patient assessment.

Artifact and noise may produce spurious data. Ensure artifact-free monitoring as much

as possible and assess the patient frequently to confirm the appropriateness of

monitor data.