User tests, Maintaining the equipment – Physio-Control LIFEPAK 15 User Manual

Page 211

Advertising
background image

MAINTAINING THE EQUIPMENT

10

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

10-5

• Transmits the self-test results if the

TRANSMIT RESULTS

option is enabled. (Transmission may take

up to 4 minutes.)

• Turns itself off

If the defibrillator detects a problem during an auto test, it annotates the fault condition on the

printed test report.

For more information about enabling the

TRANSMIT RESULTS

option, see the LIFEPAK 15 Monitor/

Defibrillator Setup Options guide provided with your device.

The automatic self-test is not performed if the defibrillator is already turned on at 03:00 or if power

is not available. If the defibrillator is manually turned on while a self-test is in progress, the self-test

is halted and the defibrillator turns on normally.

For more information, see Table 10-2 on page 10-18.

User Tests

The User Test is a functional test of the LIFEPAK 15 monitor/defibrillator. The User Test should be

performed only as a test and not while using the defibrillator during patient care. Perform the User

Test as a part of completing the daily Operator’s Checklist.

Note:

The defibrillator must be in Manual mode to perform the User Test.

To perform a User Test separate from completing the Operator’s Checklist:

1. Press

ON

to turn on the LIFEPAK 15 monitor/defibrillator.

2. Press

OPTIONS

. The Options menu appears.

3. Select

USER TEST

. The defibrillator performs the following tasks:

• Self-tests to check the device.

• Charges to 10 joules and discharges internally (this energy is not accessible at the therapy

connector).

• Prints a Pass/Fail report.

If the LIFEPAK 15 monitor/defibrillator detects a failure during the User Test, the Service LED is

illuminated and the printed report indicates that the test failed. Remove the defibrillator from use

and contact a qualified service technician.

If you must interrupt the User Test, turn the power off and then on again. The test stops and the

defibrillator operates normally. A Pass/Fail report does not print.

Note:

During the User Test, all front panel controls (except

ON

) and standard paddle controls are

disabled. Routinely testing the defibrillator consumes battery power; maintain all batteries as

described in "Battery Warnings" on page 10-12.

Note:

The last 40 User and Auto Test results are transmitted with all reports to the CODE-STAT

Suite data management system.