Removing pediatric paddles, Placing pediatric paddles, Standard paddles – Physio-Control LIFEPAK 15 User Manual

Page 162

Advertising
background image

6-8

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Standard Paddles

Note:

Do not use conductive gel between adult and pediatric paddles.

Figure 6-6 Attaching a Pediatric Paddle

Removing Pediatric Paddles

To remove pediatric paddles:

1. Press down on the rear tab.

2. Slide the pediatric paddle off (see Figure 6-7).

Figure 6-7 Removing a Pediatric Paddle

Placing Pediatric Paddles

Adult paddles are recommended if the paddles fit completely on the child’s chest. Allow at least

2.5 cm (1 in.) of space between the paddles.

For infants with very small chests, pediatric paddles may be too large to place in the anterior-

lateral position. In this situation, place paddles in the anterior-posterior position. Holding the

paddles against the chest and back supports the patient on his or her side.

Do not use the pediatric paddles on adults or older children. Delivery of recommended adult

energies through this relatively small electrode surface increases the possibility of skin burns.

Anterior-Lateral Placement

Standard pediatric paddle placement includes (see Figure 6-8):

STERNUM

paddle to the patient’s right upper torso, lateral to the sternum and below the

clavicle.

APEX

paddle lateral to the patient’s left nipple in the midaxillary line, with the center of the

paddle in the midaxillary line, if possible.