MAINTAINING THE EQUIPMENT

10

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

10-15

Cleaning the Device

Clean the LIFEPAK 15 monitor/defibrillator, therapy and ECG cables, and batteries with a damp

sponge or cloth. Use only the cleaning agents listed below:

• Quaternary ammonium compounds

• Isopropyl alcohol

• Peracetic (peroxide) acid solutions

Note:

Carefully clean the connector ports. Do not allow cleaning fluids to penetrate the exterior

surfaces of the device.

Clean the carrying case accessory as follows and as described on its instruction tag:

• Hand wash using mild soap or detergent and water. A scrub brush may be useful for heavily

soiled spots. Cleaners such as Formula 409

®

are helpful for grease, oil, and other tough stains.

For information about cleaning the reusable monitoring sensors and cables, see the individual

monitoring section.

CAUTION

POSSIBLE EQUIPMENT DAMAGE

Do not clean any part of this device or its accessories with bleach, bleach dilution, or

phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not

attempt to sterilize this device or any accessories unless otherwise specified in

accessory operating instructions.