Troubleshooting tips, Data transmission – Physio-Control LIFEPAK 15 User Manual

Page 193

Advertising
background image

DATA TRANSMISSION

8

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

8-15

Troubleshooting Tips

Table 8-3 Troubleshooting Tips for Data Transmission

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Bluetooth icon on
LIFEPAK 15 monitor has
red X across it

WIRELESS

is set to OFF in the

Bluetooth Setup menu

• Set WIRELESS to ON.
• If red X remains, Bluetooth

module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.

WIRELESS

is set to OFF in the setup

options, so the WIRELESS default
is OFF each time the LIFEPAK 15
monitor is turned on

• Change WIRELESS setup option.

See LIFEPAK 15 Monitor/
Defibrillator Setup Options
provided with your device.

• If red X remains, Bluetooth

module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.

Bluetooth module in LIFEPAK 15
monitor may be faulty

• Contact qualified service

representative.

Bluetooth LED is not
illuminated

Target device is off or cannot
communicate with the
LIFEPAK 15 monitor

• Confirm that target device is on

and discoverable.

• See the operating instructions for

your target device.

Bluetooth module in LIFEPAK 15
monitor may be faulty

• If other troubleshooting is

unsuccessful, contact qualified
service representative.

LIFEPAK 15 monitor
does not automatically
connect to last connected
device

Target device is off or cannot
communicate with the LIFEPAK
15 monitor

• Confirm that target device is on

and discoverable.

Last connection to target device
may have occurred when the
LIFEPAK 15 monitor was in a
different mode

• Confirm that OUTPUT PORT is set to

BLUETOOTH WIRELESS

.

• Select FIND DEVICES and establish

a new connection.

Device does not connect
to last connected device
after WIRELESS is set to
ON

Bluetooth menu is displayed,
which prevents discovery of
devices

• Press HOME SCREEN to exit menu

and allow LIFEPAK 15 monitor to
find last connected device.

UNABLE TO CONNECT

message appears

LIFEPAK 15 monitor cannot
establish wireless connection.
Target device may not have the
necessary software application or
cannot accept data.

• Verify target device is ready to

receive transmissions.

• Attempt to retransmit.