LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Dangers and Warnings

SHOCK OR FIRE HAZARD

Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling

any fluids on defibrillator or accessories. Spilled liquids may cause the defibrillator

and accessories to perform inaccurately or fail. Do not clean with ketones or other

flammable agents. Do not autoclave or sterilize this defibrillator or accessories unless

otherwise specified.

POSSIBLE FIRE

Use care when operating this device close to oxygen sources (such as bag-valve-mask

devices or ventilator tubing). Turn off gas source or move source away from patient

during defibrillation.

ELECTRICAL INTERFERENCE HAZARDS

POSSIBLE ELECTRICAL INTERFERENCE WITH DEVICE PERFORMANCE

Equipment operating in close proximity may emit strong electromagnetic or radio

frequency interference (RFI), which could affect the performance of this device. If use

of equipment in close proximity is necessary, observe the device to verify normal

operation in the configuration in which the device will be used. RFI may result in

distorted ECG, incorrect ECG lead status, failure to detect a shockable rhythm,

cessation of pacing, or incorrect vital sign measurements. Avoid operating the device

near cauterizers, diathermy equipment, or other portable and mobile RF

communications equipment. Do not rapidly key EMS radios on and off. Refer to

Appendix D for recommended distances of equipment. Contact Physio-Control

Technical Support if assistance is required.

POSSIBLE ELECTRICAL INTERFERENCE

Using cables, electrodes, or accessories not specified for use with this defibrillator

may result in increased emissions or immunity from electromagnetic or radio

frequency interference (RFI) which could affect the performance of this defibrillator or

of equipment in close proximity. Use only parts and accessories specified in these

operating instructions.

POSSIBLE ELECTRICAL INTERFERENCE

This defibrillator may cause electromagnetic interference (EMI) especially during

charge and energy transfers. EMI may affect the performance of equipment operating

in close proximity. Verify the effects of defibrillator discharge on other equipment prior

to using the defibrillator in an emergency situation, if possible.

IMPROPER DEVICE PERFORMANCE HAZARDS

POSSIBLE IMPROPER DEVICE PERFORMANCE

Using other manufacturers’ cables, electrodes, power adapters, or batteries may cause

the device to perform improperly and may invalidate the safety agency certifications.

Use only the accessories that are specified in these operating instructions.

WARNINGS (CONTINUED)