Data transmission – Physio-Control LIFEPAK 15 User Manual

Page 195

Advertising
background image

DATA TRANSMISSION

8

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

8-17

Unable to transmit using
a gateway device that has
a functioning direct
connection or Bluetooth
connection

Transmission sites are not set up
in LIFEPAK 15 monitor

• Define transmission sites. Each

site name must exactly match the
name of the target device. See
LIFEPAK 15 Monitor/Defibrillator
Setup Options
provided with your
device.

Transmission site names in
LIFENET System do not match
site names in LIFEPAK 15
monitor

• Check site names in LIFENET

System.

Cellular communication is not
working between the gateway and
transmission sites

• Use alternate method to

communicate patient data.

UNABLE TO TRANSMIT

message appears

The LIFEPAK 15 monitor cannot
connect to the device name
selected

• Verify target device is ready to

receive transmissions.

• Verify target device setup.
• Attempt to retransmit.

The output port on the
LIFEPAK 15 monitor is not
configured for the transmission
method you are using

• Make sure the transmission site

OUTPUT PORT

is configured for the

type of transmission you are
attempting.

• Attempt to retransmit.

Target device unable to connect or
unable to connect within timeout
interval

• Verify target device is ready to

receive transmissions.

• Verify target device setup.
• Attempt to retransmit.

The target device requires you to
“accept” incoming
communications

• Check your target device for a

required acknowledgment to
connect.

• Enter passcode, when prompted.
• Set to “Always allow” if possible.
• Attempt to retransmit.

Direct connection was disrupted

• Verify cable connections.
• Attempt to retransmit.

TRANSMISSION FAILED

message appears

Computer application program is
not ready or is not available to
receive transmission

• Verify target device is running

necessary software.

• Attempt to retransmit.

Table 8-3 Troubleshooting Tips for Data Transmission (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising