Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 57

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-13

Monitor displays dashes
(---) instead of heart rate

Heart rate is < 20 bpm

• Use ECG printout to calculate

heart rate.

Heart rate is > 300 bpm

• Use ECG printout to calculate

heart rate.

Pacing function is active

• No corrective action needed.

Poor ECG signal quality

Poor electrode-skin contact

• Reposition cable or lead wires to

prevent electrodes from pulling
away from patient.

• Secure trunk cable clasp to

patient’s clothing.

• Prepare skin and apply new

electrodes.

Outdated, corroded, or dried-
out electrodes

• Check Use By date on electrode

packages.

• Use only unexpired silver/silver

chloride electrodes. Leave
electrodes in sealed pouch until
time of use.

Loose connection.
Damaged cable or connector/
lead wire

• Check or reconnect cable

connections.

• Inspect ECG and therapy cables.

Replace if damaged.

• Check cable with simulator and

replace if malfunction observed.

Noise because of radio
frequency interference (RFI)

• Check for equipment causing RFI

(such as a radio transmitter) and
relocate or turn off equipment
power.

Baseline wander
(low frequency/high
amplitude artifact)

Inadequate skin preparation

• Prepare skin and apply new

electrodes.

Poor electrode-skin contact

• Check electrodes for proper

adhesion.

Diagnostic frequency response

• Print ECG in monitor frequency

response.

Fine baseline artifact (high
frequency/low amplitude)

Inadequate skin preparation

• Prepare skin and apply new

electrodes.

Isometric muscle tension in
arms/legs

• Confirm that limbs are resting on

a supportive surface.

• Check electrodes for proper

adhesion.

Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising