Monitoring the ecg – Physio-Control LIFEPAK 15 User Manual

Page 58

Advertising
background image

4-14

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring the ECG

For general troubleshooting tips, see Table 10-2 on page 10-18.

ECG amplitude too small

Poor electrode-skin contact

• Prepare skin and apply new

electrodes.

ECG lead selected

• Increase ECG gain or change ECG

lead.

Patient condition (for example,
significant myocardial muscle
loss or tamponade)

• Increase ECG gain or change ECG

lead.

Monitor displays dashed
lines with no ECG LEADS
OFF

messages

PADDLES

lead selected but

patient connected to ECG
cable

• Select one of the limb or

precordial leads.

Monitor shows isoelectric
(flat) line and PADDLES lead
selected

The Test Load is connected to
therapy cable

• Remove the Test Load and

connect therapy electrodes to
cable.

• Connect ECG cable and select

another lead.

Internal pacemaker pulses
difficult to see

Pacemaker pulses are very
small

• Turn on internal pacemaker

detector (see "Monitoring Patients
Who Have Internal Pacemakers"
on page 4-11).

Monitor frequency response
limits visibility

• Connect ECG cable and select a

lead other than PADDLES.

• Print ECG in Diagnostic mode

(see "How to Print a Current
Report" on page 7-10).

Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising