Dc power adapter operation, Using the power adapter – Physio-Control LIFEPAK 15 User Manual

Page 202

Advertising
background image

9-6

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Using the Power Adapter

DC Power Adapter Operation

To use the DC Power Adapter:

1. Connect the DC power cable to the power adapter and a 12 Vdc power source.

2. Verify that the green LED strip illuminates.

3. Connect the power adapter output cable to the power adapter.

4. Connect the green end of the power adapter output cable to the auxiliary power connector on the

back of the monitor/defibrillator.

5. Verify that the

AUXILIARY POWER

LED on the defibrillator is illuminated.

6. If at least one battery is installed in the device, verify that the

BATTERY CHARGING

indicator is

illuminated or flashing. Indicator behaviors are shown in Table 9-2.

7. Press the defibrillator

ON

button.

Table 9-2 Battery Charging Indicator Behaviors

INDICATOR

DESCRIPTION

Steady green

Installed batteries are fully charged.

Flashing green

One or both installed batteries are being charged.

Off

No batteries are installed, or a battery is unable to be charged.

Advertising