Therapy – Physio-Control LIFEPAK 15 User Manual

Page 145

Advertising
background image

THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-31

ABNORMAL ENERGY DELIVERY

message appears and Shock XJ
Abnormal annotated on printout

Open air discharge with
standard paddles

• Press paddles firmly on

patient’s chest when
discharging.

Standard paddles placed face-
to-face when

(shock)

button

pressed

• Perform test discharges per

Operator’s Checklist.

• See "Manual Defibrillation

Warnings" on page 5-23.

Patient impedance is out of
range

• Increase energy or repeat

shocks as needed.

• Consider replacing

disposable therapy
electrodes with new ones.

Internal fault occurred

• Repeat shock.
• Perform CPR and obtain

another defibrillator, if
necessary.

CONNECT ELECTRODES

message

appears

Therapy electrodes are not
connected to the therapy cable

• Check for electrode

connection.

Electrodes do not adhere
properly to the patient

• Press electrodes firmly on

patient’s skin.

• Clean, shave, and dry the

patient’s skin as
recommended.

• Apply new electrodes.

Electrodes are dry, damaged, or
out of date

• Apply new electrodes.

Therapy cable damaged

• Replace therapy cable and

perform daily checks per
Operator’s Checklist.

REPLACE BATTERY

prompt occurs

Both batteries are very low

• Replace one or both

batteries immediately.

• Connect to auxiliary power

using approved power
adapter.

CPR time shown (minutes/
seconds) is different than
expected

Metronome is on

• None. The metronome

adjusts the CPR time to
ensure CPR cycle ends with
compressions.

Incorrect setup option selected

• Change CPR time setup

option. See LIFEPAK 15
Monitor/Defibrillator Setup
Options
provided with your
device

.

Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising