LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

Visual Alarms

Alarms are indicated visually by:

• The violated parameter flashes in inverse video with a message in the message

area of the display.

• These visual indications remain on the display until the alarm is corrected. Visual

indication of alarms continue even when the tones have been silenced.

Table A-5 Setup Options Factory Default Settings

MENU

MENU/ITEM

FACTORY DEFAULT SETTINGS

General

Language

(Country Specific)

Code Summary

Long

Trend Summary

Off

Site Number

000

Device ID

“LP15” + last 4 digits of serial number, for example,
LP151234

Auto Log

On

Line Filter

60 Hz

Timeout Speed

30 seconds

Manual mode

Sync After Shock

Off

Pads Default

200 (joules)

Energy Protocol

Inactive

Internal Default

10 (joules)

Voice Prompts

On

Shock Tone

On

Manual Access

Manual / Direct

Set Passcode

0000

AED mode

Energy Protocol

200–300–360

Auto Analyze

Off

Motion Detection

On

Pulse Check

Never

CPR

CPR Time 1

120 seconds

CPR Time 2

120 seconds

Initial CPR

Off

Initial CPR Time

120 seconds

Preshock CPR

Off

CPR Metronome Metronome

On

Adult - No Airway

30:2

Adult - Airway

10:1

Youth - No Airway

15:2

Youth - Airway

10:1

Table A-4 Alarm Performance Characteristics (Continued)

CHARACTERISTIC DESCRIPTION