Troubleshooting tips, Therapy – Physio-Control LIFEPAK 15 User Manual

Page 143

Advertising
background image

THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-29

Troubleshooting Tips

Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Charge time to 360 joules
exceeds 10 seconds

Battery low

• Replace battery with fully

charged battery.

Operating temperature is too
low

• Move patient and device to

warmer environment, if
necessary.

Energy not delivered to patient
when

(shock)

buttons are

pressed

Device is in Sync mode and
QRS complexes are not
detected

• Adjust ECG size for

optimum sensing QRS or
deactivate SYNC if rhythm
VF/VT.

SYNC

accidentally pressed and

rhythm is VF/VT

• Press SYNC to turn off Sync.
• Press

(shock)

buttons.

Device in Sync mode and

(shock)

buttons not pressed

and held until next detected
QRS

• Hold

(shock)

buttons

until discharge occurs or
next detected QRS and
ENERGY DELIVERED

message

appears.

(shock)

buttons pressed

before full charge reached

• Wait for tone and message

indicating full charge.

Standard paddles connected
and

(shock)

button on

defibrillator front panel pressed

• Simultaneously press

(shock)

buttons on standard

paddles to discharge.

Sixty seconds elapsed before

(shock)

buttons were

pressed after full charge.
Energy was internally removed.

• Press

(shock)

buttons

within 60 seconds of full
charge.

Energy selection changed

• Press CHARGE again.

CONNECT CABLE

message

appears

Therapy cable disconnected
during charging

• Reconnect cable and press

CHARGE

again.

Therapy cable damaged

• Replace therapy cable and

perform daily checks per
Operator’s Checklist.

ENERGY FAULT

message appears

(selected and available energy)

Defibrillator out of calibration

• Attempt to transfer energy.
• Contact a qualified service

technician.