Nibp monitoring warnings and caution, How nibp monitoring works, Monitoring noninvasive blood pressure – Physio-Control LIFEPAK 15 User Manual

Page 82

Advertising
background image

4-38

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring Noninvasive Blood Pressure

NIBP Monitoring Warnings and Caution

How NIBP Monitoring Works

The NIBP monitor uses the oscillometric measurement technique. The oscillometric technique does

not use Korotkoff sounds to determine blood pressure; rather, it monitors the changes in pressure

pulses that are caused by the flow of blood through the artery. The NIBP monitor inflates the cuff

around the patient’s arm to a value that occludes the artery, and then deflates the cuff in steps.

When blood starts to flow through the artery, the increasing blood flow causes the amplitude of the

pressure pulses in the cuff to increase. As the NIBP monitor steps the pressure down, the pulses

WARNINGS

POSSIBLE LOSS OF INTRAVENOUS ACCESS AND INACCURATE INFUSION RATE

Do not apply the blood pressure cuff on an extremity that is used for an intravenous

infusion. Patency of the intravenous infusion may be affected by blood pressure

measurement due to the occlusion of blood flow.

POSSIBLE CIRCULATION IMPAIRMENT

Prolonged, continuous use of a blood pressure cuff may impair blood flow to the

extremity. Check circulation regularly and loosen or reposition the cuff if changes in

circulation occur.

INACCURATE READINGS HAZARDS

POSSIBLE INACCURATE BLOOD PRESSURE READINGS

Do not alter the NIBP monitor’s pneumatic tubing. Altering NIBP tubing may cause

improper performance and may void the warranty. Avoid compression or restriction of

pressure tubes.

POSSIBLE INACCURATE BLOOD PRESSURE READINGS

Using NIBP accessories not recommended by Physio-Control may cause the device to

perform improperly and invalidate the safety agency certifications. Use only the

accessories that are specified in these operating instructions.

POSSIBLE INACCURATE OXYGEN SATURATION READINGS

Do not perform NIBP measurement on an extremity used for oxygen saturation

monitoring. Oxygen saturation measurement is affected by blood pressure

measurement due to the occlusion of blood flow.

CAUTION

EQUIPMENT DAMAGE

Do not inflate a cuff unless it is placed on an extremity.