LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Front View

Connecting and Disconnecting the Therapy Cable

IMPORTANT!

The LIFEPAK 15 monitor/defibrillator QUIK-COMBO therapy cable and standard

(hard) paddles have the same type of connector and connect to the defibrillator at the same

location. These therapy cables are not compatible with other LIFEPAK defibrillator/monitors.

To connect a therapy cable to the defibrillator:

Figure 3-9 Connect Therapy Cable

To disconnect the therapy cable from the defibrillator:

Figure 3-10 Disconnect Therapy Cable

WARNING

POSSIBLE EQUIPMENT DAMAGE AND INABILITY TO DELIVER THERAPY

To help protect the therapy cable connector from damage or contamination, keep

therapy cable connected to the defibrillator at all times. Inspect and test the therapy

cable daily according to the Operator’s Checklist in the back of this manual. Physio-

Control recommends replacement of therapy cables every three years to reduce the

possibility of failure during patient use.

1. Align the therapy cable connector with the

receptacle.

2. Slide the therapy cable until you feel the

connector lock in place. You will also hear a

“click.”

1. Press the release button on the therapy

cable connector.

2. Slide the therapy cable connector out.