LIFEPAK 15 Monitor/Defibrillator Operating Instructions

QUIK-COMBO Therapy Electrodes

Connecting Therapy Electrodes

To connect QUIK-COMBO therapy electrodes to the QUIK-COMBO therapy cable:

1. Open the protective cover on the therapy cable connector (see Figure 6-2).

2. To insert the QUIK-COMBO electrode connector into the therapy cable connector, align the

arrows and press the connectors firmly together.

Figure 6-2 Connect QUIK-COMBO Electrodes to Therapy Cable

Replacing and Removing Therapy Electrodes

Replace adult QUIK-COMBO electrodes with new electrodes after one of the following occurs:

• 50 defibrillation shocks

• 24 hours on the patient’s skin

• 8 hours of continuous pacing

Replace pediatric QUIK-COMBO electrodes with new electrodes after one of the following occurs:

• 25 defibrillation shocks

• 24 hours on the patient’s skin

• 8 hours of continuous pacing

Table 6-1 QUIK-COMBO Electrodes

TYPE

DESCRIPTION

QUIK-COMBO

Electrodes, with 61 cm (2 ft) of lead wire, designed for patients
weighing 15 kg (33 lb) or more

QUIK-COMBO RTS

Electrodes, providing a radio-transparent electrode and lead wire set,
designed for patients weighing 15 kg (33 lb) or more

QUIK-COMBO with
REDI-PAK™ preconnect
system

Electrodes designed for patients weighing 15 kg (33 lb) or more and
that allow preconnection of the electrode set to the device while
maintaining electrode shelf life and integrity

Pediatric QUIK-COMBO RTS

Electrodes designed for patients weighing 15 kg (33 lb) or less

Therapy cable

connector

QUIK-COMBO

electrode

connector