Manual defibrillation warnings, Manual mode, Therapy – Physio-Control LIFEPAK 15 User Manual

Page 137

Advertising
background image

THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-23

Manual Defibrillation Warnings

Manual Mode

The LIFEPAK 15 monitor/defibrillator is set up to operate in Manual mode when it is turned on

(factory default setting). If required by your protocols, the defibrillator can be set up to power on in

the automated external defibrillator (AED) mode. For information on switching from AED mode to

Manual mode, see "Switching from AED Mode to Manual Mode" on page 5-14.

WARNINGS

SHOCK HAZARD

Conductive gel (wet or dry) on the paddle handles can allow the electrical energy to

discharge through the operator during defibrillation. Completely clean the paddle

electrode surfaces, handles, and storage area after defibrillation.

BURNS AND INEFFECTIVE ENERGY DELIVERY HAZARDS

POSSIBLE FIRE, BURNS, AND INEFFECTIVE ENERGY DELIVERY

Precordial lead electrodes and lead wires may interfere with the placement of standard

paddles or therapy electrodes. Before defibrillation, remove any interfering precordial

lead electrodes and lead wires.

POSSIBLE BURNS AND INEFFECTIVE ENERGY DELIVERY

A gel pathway on the skin between the standard paddles will cause defibrillating

energy to arc between paddles and divert energy away from the heart muscle. Do not

allow conductive gel (wet or dry) to become continuous between paddle sites.

POSSIBLE PATIENT SKIN BURNS

During defibrillation, air pockets between the skin and standard paddles can cause

patient skin burns. Completely cover paddle electrode surfaces with fresh conductive

gel and apply 25 lb of pressure per paddle during discharge.

POSSIBLE PADDLE DAMAGE AND PATIENT SKIN BURNS

Discharging the defibrillator with the standard paddle surfaces shorted together can

pit or damage the paddle electrode surface. Pitted or damaged paddle surfaces may

cause patient skin burns during defibrillation. Discharge the defibrillator only as

described in these operating instructions.

POSSIBLE INCORRECT ENERGY DELIVERY

The defibrillator does not automatically adjust energy when using pediatric therapy

electrodes or pediatric standard paddles. Manually select the appropriate energy prior

to defibrillating the patient.

Advertising