Appendix – Physio-Control LIFEPAK 15 User Manual

Page 243

Advertising
background image

APPENDIX

A

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

A-9

Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):

Battery Status

Charge Time

Fully charged

200 J within 15 seconds, nominal

Fully charged, followed by 15
full-energy shocks

360 J within 30 seconds

Fully charged

360 J within 30 seconds

Maximum Time from Power-on to Shock Ready (Manual Mode):

Battery Status

Charge Time

Fully charged, followed by 15
full-energy shocks

360 J within 25 seconds

Maximum Time from Power-on to Shock Ready (AED Mode):

Battery Status

Charge Time

Fully charged, followed by 15
full-energy shocks

360 J within 40 seconds

Manual Mode

Energy Select

2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 125, 150, 175,
200, 225, 250, 275, 300, 325, and 360 joules

Synchronous
cardioversion

The maximum time delay between synchronization pulse and the delivery
of energy, once the output has been activated, is not more than 60 msec.
This time delay is measured from the peak of the QRS to the peak of the
defibrillator waveform.

Paddles Lead Off
Sensing

The transition point at which device changes from assuming that QUIK-
COMBO electrodes are properly connected to patient to assuming that
electrodes are not connected is 300±50

.

Biphasic Waveform

Biphasic Truncated Exponential

The following specifications apply from 25 to 200

, unless otherwise

specified:

Energy Accuracy: ±1 joule or 10% of setting, whichever is greater, into
50

±2 joules or 15% of setting, whichever is greater, into 25-175.

Voltage Compensation: Active when disposable therapy electrodes are
attached. Energy output within ±5% or ±1 joule, whichever is greater, of
50

 value, limited to the available energy which results in the delivery of

360 joules into 50



Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)

CHARACTERISTIC

DESCRIPTION

Advertising