Therapy – Physio-Control LIFEPAK 15 User Manual

Page 135

Advertising
background image

THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-21

For general troubleshooting tips, see Table 10-2 on page 10-18.

Home Screen is blank but ON
LED is illuminated

Screen not functioning
properly

• Press ANALYZE and follow

voice prompts to treat
patient.

Analysis result is NO SHOCK
ADVISED

and ECG shows a

perfectly flat, isoelectric line.

The Test Load is connected to
therapy cable

• Remove the Test Load and

connect therapy electrodes to
the cable.

Table 5-1 Troubleshooting Tips for AED Mode (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Advertising