Table B-1 Summary of Screen Messages

MESSAGE

DESCRIPTION

12-LEAD ECG UNAVAILABLE

A 12-lead was requested but the necessary ECG data is not
available.

ABNORMAL ENERGY DELIVERY

A discharge occurred when the paddles were shorted together,
when hard paddles did not have adequate contact with the
patient or were discharged in the air, or patient impedance was
out of range. Message may also appear in certain types of
internal faults.

ACCESS DENIED

Three consecutive incorrect passcode attempts were made to
enter Manual mode.

ACQUIRING 12-LEAD

Monitor is acquiring data for 12-lead ECG report.

ACQUIRING SNAPSHOT

A snapshot report of current vital signs has been requested.

ADVISORY MODE-MONITORING

The device is monitoring the patient ECG for a shockable
rhythm.

ADVISORY: SPCO > 10%

SpCO advisory alert activated. SpCO value is greater than 10%.

ADVISORY: SPMET > 3%

SpMet advisory alert activated. SpMet value is greater than 3%.

ALARM APNEA

No valid breath has been detected for 30 seconds.

ALARMS SILENCED

Alarms are silenced. An alert tone with status message ALARMS
SILENCED

occurs periodically as a reminder.

ANALYZING 12-LEAD

The data for 12-lead ECG report is being analyzed.

ANALYZING NOW-STAND CLEAR

The AED is analyzing the patient ECG rhythm.

ATTEMPTING TO TRANSMIT

The device is processing a transmission request.

AUTO NIBP CANCELLED

The automatic initiation of NIBP measurements has been
cancelled.

BATTERY X LOW

The specified battery has a low energy condition.

BLUETOOTH DEVICE NOT FOUND

Bluetooth device has not been detected.

BLUETOOTH UNAVAILABLE

Unable to locate or connect to target device.

CANNOT CHARGE

CHARGE

is pressed and the synchronize source is missing for

synchronized cardioversion, the therapy cable is not connected,
or QUIK-COMBO electrodes are not attached to the therapy
cable.

CANNOT CHARGE BATTERIES

Both batteries are installed, and the device is unable to charge
either battery.

CANNOT CHARGE BATTERY 1

The device is unable to charge the battery in battery well 1.

CANNOT CHARGE BATTERY 2

The device is unable to charge the battery in battery well 2.