Oximeter monitoring procedure, Monitoring spo2, spco, and spmet – Physio-Control LIFEPAK 15 User Manual

Page 74

Advertising
background image

4-30

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring SpO2, SpCO, and SpMet

Oximeter Monitoring Procedure

Power to the pulse oximeter is controlled by the LIFEPAK 15 monitor/defibrillator. When the

defibrillator is turned on, the oximeter turns on and performs a calibration and self-test that

requires approximately 20 seconds. During the calibration and self-test, the screen does not display

SpO

2

, SpCO, or SpMet information.

To conserve battery power, the pulse oximeter goes into “sleep mode” when not in use. Sleep mode

is activated within 10 seconds of disconnecting the sensor. During sleep mode, the screen does not

display SpO

2

, SpCO, or SpMet information. When a sensor or patient signal is detected, the

oximeter performs a self-test and then returns to normal mode.

The pulse oximeter measures and displays SpO

2

levels between 50 and 100%. SpO

2

levels less

than 50% are displayed as <50. When SpO

2

levels are between 70 and 100%, oximeter

measurements are accurate ±3 digits. The pulse oximeter measures and displays SpCO in the range

of 0–40% with accuracy of ±3 digits. The pulse oximeter measures and displays SpMet in the range

of 0–15% with accuracy of ± 1 digit.

To monitor SpO

2

:

1. Press

ON

.

2. Connect the pulse oximeter cable to the monitor and sensor.

3. Attach the sensor to the patient.

4. Observe the pulse bar for fluctuation. Amplitude of the pulse bar indicates relative signal

quality.

5. Confirm that the SpO

2

reading appears and is stable.

6. Use the

SPEED DIAL

to adjust volume, sensitivity, and averaging time, as necessary.

To monitor SpCO or SpMet:

1. Perform Step 2 through Step 5 above.

2. Verify that an SpCO/SpMet sensor is in use. Only Rainbow sensors are capable of reading SpCO/

SpMet.

3. Encourage the patient to remain still.

4. To quickly obtain SpCO or SpMet value, press

PRINT

. If dashes (---) appear on printout instead of

values for SpCO or SpMet, allow a few more seconds for measurement to be obtained.

or

To display SpCO or SpMet:

• Use the

SPEED DIAL

to select the SpO

2

area.

• Select

PARAMETER

from menu.

• Select

SPCO

or

SPMET

. Selected value displays for 10 seconds.

Note:

SpCO and SpMet monitoring are not intended for use under patient motion or low perfusion

conditions.

Advertising