MAINTAINING THE EQUIPMENT

10

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

10-9

Therapy Cable Monitoring and Synchronized Cardioversion Check

Equipment Needed:

• LIFEPAK 15 monitor/defibrillator

• QUIK-COMBO therapy cable

• Patient ECG cable

• 3-lead or 12-lead patient simulator with QUIK-COMBO connector

• Fully charged batteries or power adapter connected to a reliable power source

To check therapy cable monitoring and synchronized cardioversion:

1. Press

ON

.

2. Connect the ECG cable to the defibrillator and to the simulator.

3. Connect the therapy cable to the simulator.

4. Turn on the simulator and select any rhythm except asystole or ventricular fibrillation.

5. Select

PADDLES

lead.

6. Confirm that the screen displays an ECG and that the

PADDLES LEADS OFF

message does not

appear.

Note:

If the screen displays dashed lines, artifact (irregular noise signals), or any warning

message, replace the therapy cable and repeat the test. If the problem cannot be corrected,

remove the defibrillator from active use and contact a qualified service representative.

7. Select Lead

II

.

8. Press

SYNC

. Confirm that the

SYNC

LED lights and the Sync mode message appears. Adjust

ECG size until sense markers appear on the QRS complexes. Confirm that the

SYNC

LED

blinks off with each detected QRS complex and that the heart rate is displayed.

9. Select

50 JOULES

.

10.Press

CHARGE

.

CAUTION

POSSIBLE SIMULATOR DAMAGE

Do not discharge more than 30 shocks within an hour, or 10 shocks within a five-

minute period, or pace continually into Physio-Control patient simulators. Simulators

may overheat.