Etco2 monitoring procedure, Monitoring etco2 – Physio-Control LIFEPAK 15 User Manual

Page 92

Advertising
background image

4-48

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring ETCO2

Similar to the diminished slope of the Expiratory Upstroke, this pattern can occur in asthma,

chronic obstructive pulmonary disease (COPD), partial upper-airway obstruction, or partial

mechanical obstruction such as a partially kinked ETT.

Inspiratory Downstroke

The fall to baseline (IV-V segment) is a nearly vertical drop. This slope

can be prolonged and can blend with the expiratory plateau in cases of leakage in the exhale

portion of the breathing circuit. The peak EtCO

2

value (IV) is often not reached. Relying on the

numeric end-tidal value without observing the breathing waveform may obscure the presence of

a leak.

EtCO2 Monitoring Procedure

When activated, the EtCO

2

monitor draws power from the defibrillator. The LIFEPAK 15 monitor/

defibrillator activates the EtCO

2

monitor when it senses the attachment of the FilterLine set.

Initialization, self-test, and warm up of the EtCO

2

monitor is typically less than 30 seconds, but

may take up to two-and-one-half minutes.

To monitor EtCO

2

:

1. Press

ON

.

2. Select the appropriate EtCO

2

accessory for the patient.

3. Open the CO

2

port door and insert the FilterLine connector; turn connector clockwise until tight.

4. Verify that the CO

2

area is displayed. The EtCO

2

monitor performs the autozero routine as part

of the initialization self-test.

Note:

If you use a ventilation system, do not connect the FilterLine set to the patient/ventilation

system until the EtCO

2

monitor has completed its self-test and warm-up.

5. Display CO

2

waveform in Channel 2 or 3.

6. Connect the CO

2

FilterLine set to the patient.

7. Confirm that the EtCO

2

value and waveform are displayed. The monitor automatically selects the

scale for the best visualization of the waveform. You can change the scale, if desired, as

described in the next section.

Note:

It is possible for the FilterLine set to become loose at the device connection and still have an

EtCO

2

value and CO

2

waveform, but they may be erroneously low. Make sure the FilterLine

connection is firmly seated and tight.

CAUTION

POSSIBLE EQUIPMENT DAMAGE

Failure to replace a broken or missing CO

2

port door may allow water or particulate

contamination of the internal CO

2

sensor. This may cause the CO

2

module to

malfunction.