BASIC ORIENTATION

3

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

3-23

When alarms are ON, you can silence them preemptively for up to 15 minutes.

To silence alarms preemptively:

1. Press

ALARMS

.

2. Select

SILENCE

.

3. Select

SILENCE

duration of 2, 5, 10, or 15 minutes.

The message

ALARMS SILENCED

appears in the message area at the bottom of the Home Screen.

Note:

When you select

SILENCE

, the VF/VT alarm is not silenced.

Options

Press

OPTIONS

to display the Options menu. Rotate the

SPEED DIAL

to scroll through the choices.

Press the

SPEED DIAL

to make a selection.

Table 3-9 Options Menu Selections

SELECTION

DESCRIPTION

FOR MORE INFORMATION

Patient

Enter patient name, patient ID, incident, age, and
sex.

See "Entering Patient
Data" in next
section

Pacing

Select demand or nondemand pacing. Set internal
pacer detection ON or OFF.

See page 5-32

Date/Time

Set date and time. Cycle power for change to take
effect.

See LIFEPAK 15
Monitor/Defibrillator
Setup Options
for time
display options.

Alarm Volume

Adjust volume for alarms, tones, voice prompts and
CPR metronome.

Archives

Access archived patient records.

See page 7-11

Options

Patient...

Pacing...

Date/Time...

Alarm Volume...

Archives...

Print...

User Test