Troubleshooting tips, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 95

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-51

Troubleshooting Tips

Table 4-6 Troubleshooting Tips for EtCO

2

Monitoring

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

ALARM

APNEA message appears

and waveform is solid line at or
near zero

No breath has been detected
for 30 seconds since last valid
breath

• Check the patient.

FilterLine connection to device
is loose

• Twist FilterLine connector

clockwise until tight and
firmly seated.

FilterLine set is disconnected
from patient or ETT

• Check ventilation

equipment (if used) for
leaks or disconnected
tubing.

CO2 FILTERLINE OFF

message

appears and waveform is “---”

FilterLine set disconnected or
not securely connected to
device

• Connect FilterLine set to

device port.

• Twist FilterLine connector

clockwise until tight and
firmly seated.

CO2 FILTERLINE PURGING

message appears and waveform
is “---”

FilterLine set is kinked or
clogged with fluid, or rapid
altitude change occurred

• Disconnect and then

reconnect the FilterLine set.

• Twist FilterLine connector

clockwise until tight and
firmly seated.

CO2 FILTERLINE BLOCKAGE

message appears and waveform
is “---”

The message appears after 30
seconds of unsuccessful
purging

• Disconnect and then

reconnect the FilterLine set.

• Change the FilterLine set.

FilterLine set is kinked or
clogged

• Twist FilterLine connector

clockwise until tight and
firmly seated.

CO2 INITIALIZING

message

appears and waveform is “---”

FilterLine set connected to
device while module is
initializing

• None.

Defibrillation shock delivered

• None. System resets

automatically within 20
seconds.

AUTO ZEROING

message appears

and waveform is “---”

Module is performing self-
maintenance

• None.

Defibrillation shock delivered

• None. System resets

automatically within 20
seconds.