Printed 12-lead ecg frequency response, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 65

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-21

Note:

The fiducial marks displayed in the 4-channel format identify the measurement intervals

used for the interpretive statements of the analysis program. These marks are part of the

analysis program and cannot be turned off.

Figure 4-9 Example of Printed 4-Channel, Standard 12-Lead ECG Report

Figure 4-10 Example of Printed 4-Channel, Cabrera 12-Lead ECG Report

Printed 12-Lead ECG Frequency Response

The 12-lead ECG can be printed in two diagnostic frequency responses (or bandwidths):

0.05–40 Hz and 0.05–150 Hz. The frequency response of 0.05–150 Hz is the Association for the

Advancement of Medical Instrumentation (AAMI) standard for diagnostic ECGs. The 0.05–40 Hz

setting preserves the low frequency limit that is needed for the diagnosis of myocardial ischemia

and infarction while reducing high frequency artifact (in particular from patient muscle tension) to

help make the diagnostic printout less noisy and more readable.

I
II
III

V1
V2
V3
V4
V5
V6

0.00 mm
0.00 mm

-1.10 mm

0.00 mm
0.00 mm
0.00 mm

0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm

Fiducial marks

Record ID:
Patient ID:
Incident:
Age: 50

Sex: M

12-Lead 1

P-QRS-T Axes

24 Apr 08
PR .168s
QT/QTc

HR 78 bpm

QRS 0.104s

Abnormal ECG **Unconfirmed**
***Meets ST ELEVATION MI

Sinus rhythm
Anteroseptal ST elevation, CONSIDER ACUTE INFARCT

aVR

aVL

aVF

V1

V2

V3

V4

V5

V6

Lee, William

Name:

STJ Level

aVR
aVL
aVF



I
II
III

V1
V2
V3
V4
V5
V6

0.00 mm
0.00 mm

-1.10 mm

0.00 mm
0.00 mm
0.00 mm

0.00 mm
1.80 mm
2.50 mm
2.76 mm
0.00 mm
0.00 mm

Record ID:
Patient ID:
Incident:
Age: 50

Sex: M

12-Lead 1
24 Apr 08
PR .168s
QT/QTc
P-QRS-T Axes

Abnormal ECG **Unconfirmed**
***Meets ST ELEVATION MI CRITERIA***

Sinus rhythm
Anteroseptal ST elevation, CONSIDER ACUTE INFARCT

Name:

Lee, William

HR 89 bpm

QRS 0.104s

aVL

-aVR

aVF

V1

V2

V3

V4

V5

V6

STJ Level

aVR
aVL
aVF

Advertising