Troubleshooting tips, Monitoring the ecg – Physio-Control LIFEPAK 15 User Manual

Page 56

Advertising
background image

4-12

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring the ECG

Troubleshooting Tips

If problems occur while monitoring the ECG, check Table 4-2 for aid in troubleshooting. For basic

troubleshooting problems, such as no power, see "General Troubleshooting Tips" on page 10-18.

Table 4-2 Troubleshooting Tips for ECG Monitoring

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Any of these messages
displayed:
CONNECT ELECTRODES
CONNECT ECG LEADS
ECG LEADS OFF
XX LEADS OFF

Therapy electrodes not
connected

• Connect therapy electrode.

One or more ECG electrodes
disconnected

• Connect ECG electrode.

ECG cable is not connected to
monitor

• Connect ECG cable.

Poor electrode-skin contact

• Reposition cable or lead wires to

prevent electrodes from pulling
away from patient.

• Secure trunk cable clasp to

patient’s clothing.

• Prepare skin and apply new

electrodes.

PACER

was pressed. The

monitor automatically
switched to Lead II, but ECG
leads are not connected.

• Connect ECG leads and initiate

pacing.

Broken ECG cable lead wire

• Select another lead.
• Select PADDLES lead, and use

standard paddles or therapy
electrodes for ECG monitoring.

• Check ECG cable continuity.

Screen blank and ON LED
illuminated

Screen not functioning
properly

• Print ECG on recorder as backup.
• Contact service personnel for

repair.

Systole beeps not heard or
do not occur with each
QRS complex

Volume too low

• Adjust volume.

QRS amplitude too small to
detect

• Adjust ECG size.

Displayed heart rate (HR)
different than pulse rate

ECG size set too high or too
low

• Adjust ECG size up or down.

Monitor detecting the patient’s
internal pacemaker pulses

• Change monitor lead to reduce

internal pacemaker pulse size.

Displayed heart rate (HR)
different from displayed
ECG waveform

ECG size set too high or too
low

• Adjust ECG size up or down.

Monitor detecting the patient’s
internal pacemaker pulses

• Change monitor lead to reduce

internal pacemaker pulse size.