Appendix – Physio-Control LIFEPAK 15 User Manual

Page 259

Advertising
background image

APPENDIX

B

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

B-5

PUSH ANALYZE

Press ANALYZE to begin ECG analysis.

PUSH AND HOLD SHOCK BUTTON!

The defibrillator is in Sync mode, fully charged, and ready to
provide therapy.

PUSH AND HOLD PADDLE
BUTTONS TO SHOCK!

The defibrillator is in Sync mode, fully charged, and ready to
provide therapy with hard paddles connected.

PUSH SHOCK BUTTON!

The defibrillator is fully charged and ready to provide therapy.

PX NOT ZEROED

Transducer is connected or reconnected and is not zeroed.

PX TRANSDUCER NOT DETECTED

IP transducer is disconnected from the monitor/defibrillator.

PX ZERO FAILED

The device was unable to zero the pressure transducer.

PX ZEROED

Transducer successfully zeroed.

PX ZEROING

Monitor is establishing a zero reference.

RA LEADS OFF

ECG electrode “RA” is disconnected.

REPLACE BATTERY X

Power loss for the battery in well X is imminent.

SEARCHING FOR DEVICES

Device is attempting to identify available Bluetooth devices.

SELECT BIPHASIC ENERGY / XXX J

ENERGY SELECT

was pressed on front panel or on standard

paddles.

SELF TEST FAILED

Device detected internal error; remove device from service.

SELF TEST FAILED.
TRANSMITTING...

Device detected internal error and is transmitting test results.
Remove device from service after transmission is complete.

SELF TEST IN PROGRESS

Device is performing a self test after turning on.

SELF TEST PASSED

Device passed internal test and is available for use.

SELF TEST PASSED.
TRANSMITTING...

Device passed internal test and is transmitting test results.

SHOCK ADVISED!

The defibrillator has analyzed the patient ECG rhythm and
detected a shockable ECG rhythm.

SPCO: POOR QUALITY SIGNAL

Device is not receiving sufficient input from sensor.

SPMET: POOR QUALITY SIGNAL

Device is not receiving sufficient input from sensor.

SPO2: CHECK SENSOR

The SpO

2

sensor connection to device or application to patient

needs checked.

SPO2: LOW PERFUSION

Patient has a weak pulse.

SPO2: NO SENSOR DETECTED

A sensor is disconnected from the monitor.

SPO2: POOR QUALITY SIGNAL

Device is not receiving sufficient input from sensor.

Table B-1 Summary of Screen Messages (Continued)

MESSAGE

DESCRIPTION

Advertising