LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring SpO2, SpCO, and SpMet

Note:

Most Rainbow sensor messages (SpO

2

, SpCO, and SpMet) are reported as

SPO2: (MESSAGE)

.

The

POOR QUALITY SIGNAL

message indicates the specific parameter affected.

For general troubleshooting tips, see Table 10-2 on page 10-18.

SPCO/SPMET: POOR QUALITY
SIGNAL

message appears

When the signal quality is low,
the accuracy of the
measurement may be
compromised

• Check that sensor and cable

are connected properly.

• Move sensor to a better

perfused site.

SPO2: SENSOR DOES NOT SUPPORT
SPCO OR SPMET

message

appears

SpO

2

-only sensor used with

SpCO/SpMet capable device

• None necessary, or use

Rainbow sensor to measure
SpCO or SpMet.

Table 4-4 Troubleshooting Tips for SpO

2

, SpCO, and SpMet (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION