MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-11

Monitoring Patients Who Have Internal Pacemakers

The LIFEPAK 15 monitor/defibrillator internal pacemaker detection feature can be used to help

identify internal pacemaker pulses on the printed ECG. When enabled, this feature uses lead V4

to detect internal pacemaker pulses. If V4 is not available because it is not attached or is too

noisy, Lead II or Paddles Lead is used.

When the internal pacemaker detection feature is ON, the LIFEPAK 15 monitor/defibrillator

annotates a hollow arrow

on the printed ECG if internal pacemaker pulses are detected.

Patient history and other ECG waveform data, such as wide QRS complexes, should be used to

verify the presence of an internal pacemaker. False annotations of this arrow may occur if ECG

artifacts mimic internal pacemaker pulses. If false annotations occur frequently, deactivate the

detection feature using the

OPTIONS / PACING / INTERNAL PACER

menu (see "Options" on page 3-23).

The LIFEPAK 15 monitor/defibrillator typically does not use internal pacemaker pulses to

calculate the heart rate. However, when using therapy electrodes or standard paddles to monitor

in

PADDLES

lead, the monitor may detect internal pacemaker pulses as QRS complexes, resulting

in an inaccurate heart rate.

Large amplitude pacemaker pulses may overload the QRS complex detector circuitry so that no

paced QRS complexes are counted. To help minimize ECG pickup of large unipolar pacemaker

pulses, place ECG electrodes so the line between the positive and negative electrodes is

perpendicular to the line between the pacemaker generator and the heart.

Smaller amplitude internal pacemaker pulses may not be distinguished clearly in

PADDLES

lead.

For improved detection and visibility of internal pacemaker pulses, turn on the internal

pacemaker detector function using the

OPTIONS / PACING / INTERNAL PACER

menu or connect the

ECG cable, select an ECG lead, and print the ECG in diagnostic frequency response. For

information about configuring internal pacemaker detection, see the Pacing Setup menu in the

LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.