LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring Noninvasive Blood Pressure

Cuff Selection

The use of properly designed and sized cuffs is essential for the accurate measurement of blood

pressure. The cuff must fit snugly around the extremity to occlude the artery. For a list of BP cuffs

that are intended for use with the LIFEPAK 15 monitor/defibrillator, see the LIFEPAK 15 Monitor/

Defibrillator Accessories Catalog at store.physio-control.com.

NIBP Monitoring Procedure

The NIBP monitor inflates an occluding cuff and determines systolic and diastolic pressures, mean

arterial pressure (MAP), and pulse rate. Pressure measurements are reported in mmHg and pulse

rate in beats per minute (bpm).

Both single-measurement and specified-interval (timer-controlled) methods of blood pressure

reading are available.

The NIBP monitor draws power from the defibrillator. When the defibrillator is turned on, the NIBP

monitor conducts a self-test that takes approximately three seconds.

IMPORTANT!

The LIFEPAK 15 monitor NIBP port and tubing are not compatible or

interchangeable with the NIBP tubing that is used with other LIFEPAK monitor/defibrillators.

Changing the Initial Inflation Pressure

The initial cuff pressure should be set approximately 30 mmHg higher than the patient’s

anticipated systolic pressure. The factory default initial inflation pressure for the first

measurement is 160 mmHg. For pediatric patients, the initial cuff pressure may need to be

lowered. Initial inflation settings are 80, 100, 120, 140, 160, or 180 mmHg.

Caution should be taken not to lower the initial pressure below the adult patient’s systolic

measurement. Doing so may cause the cuff to reinflate and cause patient discomfort. For

subsequent measurements, the monitor inflates approximately 30 mmHg higher than the

previously determined systolic pressure.

To select an initial pressure:

1. Rotate the SPEED DIAL to outline the NIBP

area.

2. Press the SPEED DIAL. The NIBP menu

appears.

3. Select INITIAL PRESSURE.

4. Rotate the SPEED DIAL to the desired

pressure.

5. Press the SPEED DIAL to set the initial

pressure.

NIBP

Start

Interval

Initial Pressure

Off

160 mmHg